Mains supply: 380/220/110 V, 600 W
Pump: Low-noise, compound-turbine blower, type SV 5.130/2; 40 m3/h air flow rate
Filter paper: Glass-fiber, S-class No. 8 Schleicher & Schüll; 200 mm diameter
Detector: Dual (LB150R) or triple (LB150DR) Large-area proportional counter tube, counting gas P10 or methane, gas consumption approx. 2l/h; with LB150DR, 2 cm lead, 4 p shielding (5cm on request).
Data acquisition: LB5310 P for LB150R (LB1001 optional) LB1001 for LB150DR (LB5310 optional)
Efficiency for area sources: 241AM Alpha 27% 204TI Beta 24% (with LB150DR)
Minimum detectable activities LB150R: 50 mBq/m3 (1.3 x 10-12 µCi/ml) artificial alphas within 1 h at a radon level of 10 Bq/m3 (2.7 x 10-10 µCi/ml) at a confidence level of 3-sigma
Minimum detectable activities LB150DR: 130 mBq/m3 (1.3 x 10-12 µCi/ml) artificial alphas 300 mBq/m3 (1.3 x 10-12 µCi/ml) artificial betas within 1 h at a radon level of 10 Bq/m3 (2.7 x 10-10 µCi/ml) at a confidence level of 3-sigma
Dimensions: LB150R: 55 cm wide x 69 cm deep x 124 cm high LB150DR: 55 cm wide x 60 cm deep x 160 cm high
Assembly: LB150R: 19” cabinet 20 HE LB150DR: 19” cabinet 32 HE (Mobile units available)
|