Radiation Detector: Large area scintillator (ZnS:Ag) with PMT Sensitive Area: 171 cm² Entrance Window: 6 µg aluminized plastic (0.4 mg/cm²) Grid transmission: 80%
Measuring Mode: simultaneous and separate alpha/beta measurement
Efficiencies: C-14 (Beta channel): 11 % Cl-36 (Beta channel): 56 % Tc-99m (Beta channel): 9,5 % Am-241 (Alpha channel): 21%
Background (Alpha channel): approx. 0.1 cps Background (Beta channel): approx. 10 cps Measuring Range: 0 to 5000 cps (Alpha channel) 0 to 50000 cps (Beta channel)
Spill over: Alpha- in Beta channel (Po-210): < 20% Beta- in Alpha channel (Sr-90): < 2E-5
Position dependent response to a point source: max. ± 20% (Alpha channel) max. ± 20% (Beta channel)
General specifications: Temperature range: -20°C to +40°C Humidity: 0% to 95% (without condensation) Protection: IP 53 Maximum operation with rechargeable batteries (without illumination): > 25 h Maximum operation with batteries (without illumination): > 65 h
Weight: 1300 g with batteris
Selectable calibration factors: more than 50 nuclides
Data memory: more than 1000 measured values Data output: serial port RS232
|