Radiation Protection
Home
Contact
Products
Applications
Technical Data LB 123 D-H10 Gamma Dose Rate Monitor

Basic Unit LB 1230

Display:
High-contrast dot-matrix with 32 x 84 pixels
Background illumination can be switched off
4-digits numerical values with outoranging

Control:
Membrane keypad comprising 1 on/off button and 4 soft keys

Detector connection:
8-pin connector socket Fischer 04,
spiral cable LB 75576 as standard

Data output (Option):

FSMA connector, via beam wave guide and beam wave guide interface LB 75306 with D 25 connector

Temperature range:
-15°C to 50°C

Dimensions:
145 mm (H) x 170 mm (W) x 45 mm (D)

Weight:
approx. 800 g (with batteries)

Power supply:
3 x IEC-R14 (baby cell) or 3 x rechargeable cells Varta NiCd #5014 ´

Operating life:

with R14 > 150 hours


Wall Bracket LB 1250

Low voltage supply:

+ 5 V via mains adapter LB 7619

Dimensions: 68 mm (h) x 245 mm (W) x 77 mm (D)


Transport Case

Dimensions:
180 mm (h) x 450 mm (W) x 350 mm (D)


Dose Rate Probe LB 1236-H10


Dimensions:
50 mm(Ø), 275 mm (L)
Weight: 0,46 kg
Temperature range: -10°C to 60°C
Counter tube: LB 6006 A
Calibration factor: 0,20 µSv/h per 1 cps
Measuring range: 0,05 µSv/h-10 mSv/h
Energy range: 30 keV – 2 MeV
                      (+/- 30% relative to 137Cs)



© 2008 BERTHOLD TECHNOLOGIES GmbH & Co. KG legal notices Legal Notices      print  Print Version